top of page

Talos-1

Talos 1 brings accurate, reusable, non-invasive, fluid-flow measurements without penetrating the sterile fluid boundary. Available as OEM technology or stand-alone instrumentation.

​Low flow rate infusion systems are subject to startup delays, stiction, occlusion, and environmental sensitivity. All of these problems can be addressed with a suitable flow monitoring sensor. The Talos-1 is the first real-time sensor that monitors flow rates down to 0.05 mL/h without breaking the sterile barrier or coming into physical contact with intravenous fluids.  The Talos-1 sensor clamps around an IV tube, is reusable, and monitors flow within the patient catheter while detecting flow occlusions and flow reversals. Talos-1 is a solid state device that requires no maintenance.

Talos-1 base unit for stand alone operation

Technical Specifications

​

Measurement flow rates  : 

(0.05 – 50) mL/h using Tygon ND100-80-04 microbore tubing

​

Tubing compatibility :

Sensors can be adapted to any Tygon IV tubing

​

Measurement accuracy :

10% at 25 °C, 25% across full temperature range

​

Time for measurement:  1 minute

​

Temperature Range : (15 – 35) °C

​

Sensor Dimensions : 75 mm x 28 mm

​

Sensor Weight : 120 g

​

1

1 - Other flow rates possible with different tubing sizes.

      The measurement technology works to > 50 mL/h.

​2 - All tests performed using Saint-Gobain medical Tygon tubing.

Talos-1 sensor clamp shown with microbore IV set
Talos-1 flow data compared to calibrated syringe pump

2

Talos-1 base unit, for stand-alone operation and monitoring

Talos-1 sensor clamp. Shown with 4 inch microbore IV extension set

Talus-1 flow (mL/hr)

Syringe pump flow (mL/hr)

Measured performance of Talos-1 calibrated against a 5 mL ground glass syringe driven with a New Era Pump Systems Inc. syringe pump, model number 300.